The Model 3110 represents the new generation of portable trace and percent oxygen analysis capabilities. It combines a rugged, portable analyzer with the high reliability of trace and percent level sensors. This ensures highly accurate ppm/ percent oxygen readings in a variety of background gases (including hydrocarbons). In addition, the microprocessor-based electronics provide accurate, high resolution readings and easy-to-use features.
Features:
- Ranges: 0-10 ppm to 0-10,000 ppm with 0-25% over-range
- Analyzer is intrinsically safe, FM-approved for Class I Div 1 Groups D and ATEX-approved for Zone 0 (ia) IIC*
- Sensor: Micro-fuel Cell Class B-2C
- Calibration: In ambient air or with a certified span gas
- System power requirements: AC power for battery recharge circuit of two current limited rechargeable NiCad batteries,
- 0-1V analog output
- Digital meter readout
- Steel quick-disconnect fittings with nylon cell block; mating set (3) of male fittings (vent, sample, calibration); calibration hose
- Carrying case (P/N: E834) - black, durable plastic, foam interior, or (P/N: E836) larger case for a unit with a sample system
Application:
- Air separation and liquefaction
- Pure gaseous hydrocarbon stream monitoring
- Semiconductor manufacturing
- Protective atmosphere blanketing of primary liquid feedstocks and flammable liquids
- Process analysis of gaseous monomers - vinyl chloride, propylene, butadiene, isoprene or ethylene
- Gas purity certification
- Glove box or pipeline leak detection
- Natural gas treatment and transmission
- Catalyst protection
- Inert gas welding of exotic metals
- Wave and re-flow soldering
- Heat treating and bright annealing
- Nuclear fuel processing and isotope separation
- Analysis of chemical reactions
- Head space gas analysis
- Crystal growth
- Plastics manufacturing